Brotie Co Analyzer for Gas Tester

Product Details
Customization: Available
Customized: Non-Customized
Portable: Portable
Diamond Member Since 2011

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Registered Capital
10000000 RMB
Management System Certification
ISO 9001, ISO 9000
  • Brotie Co Analyzer for Gas Tester
Find Similar Products

Basic Info.

Model NO.
261
Type
Electrochemical Gas Analyzer
Testing Principle
Ndir Testing
Range
0-2%; 0-100% CO2
Accuracy
2% Fs
Stability
1% Fs
Response Time
in 15 Second
Warranty
12 Months Analyzer; 12 Months Sensor
Transport Package
Carton
Specification
ISO9001
Trademark
BROTIE
Origin
China
HS Code
8414809090
Production Capacity
10000sets/Year

Product Description

BROTIE CO gas analyzer


Based on the NDIR principle and adopts digital precession technology as zero point/range point calibration, linear processing, temperature compensation, switch output, flow/voltage output, parameters setting. Such as digital pulsed infrared source, high sensitive infrared  sensor. It can be used for detection for different concentration and different gases as    (SO2, NOX, CO2, CO, CH4 etc).
 
Features                                                                                                   
NDIR, sensor does not touch tested gas, sensor poisoning will not happen.
Good intelligence, stability, high reliability and long calibration cycle..
Measuring cell with gilded mirror can be corrosive-resistance.  
Large LCD display.
4-20mA/0-20mA/0-10V output optional.
Two-way single-pole single throw alarm function.
Menu lock function.
Latest touch key technology.
Timing automatic storage function, which you can check store data at will;
 
Technical Specifications                                                                                       
Testing Principle:                           NDIR testing
Range:                                             0-2%; 0-100% CO
Accuracy:                                          ≤±2% FS
Stability:                                            ≤±1% FS
Response time:                               T90≤15s (500ml/min)
Repeatability:                                    ≤±1%
Sample gas pressure:                   95-105KPa
Sample gas flow:                            0.5-1.5L/min
Communication:                              RS485/RS232(Optional)
Power supply:                                  85-264VAC, 50/60Hz
Expected sensor life:                      >18000hours (in normal condition)
Expected analyzer life:                    >50000hours (in normal condition)
Weight:                                              3kg
Appearance dimension:                 144mm×144mm×295mm (W×H×D)
Measuring Medium:               Ar, O2, H2 and gases have no overlapping absorption to CO

 
Applications                                                                                        
It can be used to measure the CO concentration in Air production and manufacturing, gas detection in petrochemical process, marsh gas and industrial process.
 
 
 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.