• Non Contact LED Digita Infrared Thermometer (CE)
  • Non Contact LED Digita Infrared Thermometer (CE)
  • Non Contact LED Digita Infrared Thermometer (CE)
  • Non Contact LED Digita Infrared Thermometer (CE)
  • Non Contact LED Digita Infrared Thermometer (CE)
  • Non Contact LED Digita Infrared Thermometer (CE)

Non Contact LED Digita Infrared Thermometer (CE)

Cold Hot Dressing Application: Heat Stroke, Fever, Headache
Usage Mode: For External Use
Nature: No
Customized: Customized
Type: Thermometer
Group: All
Samples:
US$ 50/Piece 1 Piece(Min.Order)
| Request Sample
Customization:
Diamond Member Since 2011

Suppliers with verified business licenses

Manufacturer/Factory

Basic Info.

Certification
CE, SGS, ISO
Product Weight
100g(Without Battery)
Measuring Range
32-43ºC
Measuring Accuracy
±0.2ºC
Resolution
0.1ºC
Measuring Distance
1-3cm
Measuring Time
0.5seconds
Power Requirement
DV3V
Safety Standard
GB9706.1-2007
Transport Package
Carton
Origin
China
Production Capacity
3000/Day

Product Description

Non Contact LED Digita Infrared Thermometer (CE)
(Accurate    Fast    Portable)

Product weight:        100G(without battery)
Measuring range:       32-43ºC
Measuring accuracy:    ±0.2ºC
Resolution:                  0.1ºC
Measuring distance:     1-3CM
Measuring time:           0.5Seconds
Power requirement:     DV3V
Safety Standard:         GB9706.1-2007


1 Product Overview                                                
Read this product overview for a comprehensive knowledge of the medical infrared thermometers.
Read the device instructions for device status information.
Read the operation and usage to master the operation method.
Read Repair and Maintenance to learn about product repair and maintenance.
    1. Medical infrared thermometer overview
      Product name: Medical infrared thermometer
      Specifications: HPRT101, HPRT102, HPRT103
      Structural composition: The medical infrared thermometer is mainly composed of an infrared temperature sensor, a display unit, a power supply circuit, a measurement circuit and a casing.
      Scope of application: The body temperature of the subject is measured by detecting the infrared energy of the forehead of people of all ages.
    2.  Contraindications
      People with damaged skin at the target testing area
1.3  Working conditions
Ambient temperature: 10ºC
-40ºC
Relative humidity: 85% or less, non-condensing
Atmospheric pressure: 70kPa-106kPa
Warm-up time: Place it for one hour before using it to reach room temperature if the temperature during transportation or storage is
±20ºCbeyond the normal working temperature.


1.4  Product performance
1) Measuring range: 32ºC
-43ºC
2) Resolution: 0.1ºC
3) Maximum allowable error margin:
32.0
ºC-34.9ºC Maximum allowable error ±0.3ºC
35.0ºC-42.0ºCMaximum allowable error ±0.2ºC
42.1ºC-43.0ºC Maximum allowable error ±0.3ºC
4) Repeat-ability: Repeatable error S is <= 0.3ºC


1.5  Medical electrical equipment safety classification
1) Classified by type of protection against electric shock: internal power supply
2) Classified by the degree of protection against electric shock: Type B
3) Classified by the degree of protection against liquid ingress: ordinary equipment IPX0
4) Classified by the degree of safety when using flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide: non-AP / APG equipment
5) Classified by operation mode: continuous operation
6) Classified by the rated voltage and frequency: DC 3V
7) Classified by input current: Not applicable
8) Does the device have an application part that protects against defibrillation discharge effects: No
9) Does the device have a signal output or input section: No
10) Permanently installed equipment or non-permanently installed equipment: No


1.6  Graphic and symbol description
Indicates that the device type is type B
Indicates to refer to random files


2  Installation and usage instructions                              

2.1 Installing and replacing the battery
A) Open the battery cover;
B) Insert two 1.5V alkaline batteries, and make sure that the positive and negative poles correspond to the poles of the battery holder.
C) Close the battery cover


Note: Please replace the battery in time when the thermometer shows low battery.
Note:Please remove the battery when the thermometer is not used for a long time.


2.2Operation and use
2.2.1 Start
Press the measurement button to turn on.
2.2.2 Settings
HPRT101: Press the mode setting button to select the mode, which includes forehead temperature measurement (face image) and object measurement; in forehead temperature measurement mode, the display unit shows the face portrait; the default is the forehead temperature measurement mode when the device is turned on;
HPRT102: Long press the measurement button to enter the mode selection;
HPRT103: Press the mode setting button to select the mode, which includes forehead temperature measurement and object measurement. The mode status indication is as follows (the light is on, which means that the mode is set);



Object measurement
Forehead measurement                             Power indicator

2.3.3  Measurement
In the forehead temperature mode, point the thermometer at the measurement area (1-3 cm distance), press the measurement button, and a beep sound indicates that the measurement is complete.

2.3.4  Power off
It will be automatically power off when it is not in use for a long time.


3  Repair and maintenance                                          
3.1 Cleaning and disinfection
To avoid long-term damage to our products, we recommend cleaning and disinfecting them only when it is necessary. We recommend using a cotton ball dipped in 75% medical alcohol to wipe the surface of the medical infrared thermometer. The company will provide circuit diagrams and other information at the user's request to help users to have appropriate and qualified technicians repair those products that the company classifies as user-serviceable.
3.2 Storage and transportation
A) Ambient temperature: -20
ºC-55ºC
B) Relative humidity: ≤ 85%, non-condensing
C) Atmospheric pressure: 70kPa-106kPa

Non Contact LED Digita Infrared Thermometer (CE)
Non Contact LED Digita Infrared Thermometer (CE)
Non Contact LED Digita Infrared Thermometer (CE)

Non Contact LED Digita Infrared Thermometer (CE)
Non Contact LED Digita Infrared Thermometer (CE)




 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Diamond Member Since 2011

Suppliers with verified business licenses

Manufacturer/Factory
Registered Capital
10000000 RMB
Management System Certification
ISO 9001, ISO 9000