Basic Info.

Model NO.
dp
Gas
H2O
Type
Electrochemical Gas Analyzer
Range
+20oc~-60oc
Standard Type
+20oc~-80oc
Sensitivity
0.1oc
Sample Gas Flow Speed
1-3L/Min
Signal Outputs
4-20mA or 0-5V DC Isolated
Working Temperature
0oc-+45oc
Response Time
in 5min
Transport Package
Carton
Specification
ISO9001
Trademark
BROTIE
Origin
China
HS Code
8414809090
Production Capacity
10000sets/Year

Product Description

Dew point Analyzer

Features:
l         Imported capacitive dew point sensor, fast response and high sensitivity;
l         Precision filter;
l         128 × 64 dot-matrix LCD, concise menu;
l         Showing dew point, and the volume ratio (PPM/V) synchronously; intuitive readings;
l         Automatic data storage, free setting record time interval. Data can be recorded curve or tabular form and browses easily;
l         Upper and lower alarm values can be set freely during the measure ranges;
l         standard RS232 or RS485 interface;

Industries Served:
It can be used in the fields of Petrochemicals, natural gas, industrial gases, the semiconductor, drying industry, the food industry, power industry, machinery manufacturing, air separation industry, the pharmaceutical industry, and so on.

Specifications:
Range:                               Basic type: +20ºC~-60ºC
Standard type:                  +20ºC~-80ºC
Expanding type:               +20ºC~-100ºC
Accuracy:                           ±1% F.S
Response time:               T90≤5min
Sensitivity:                          0.1ºC
Stability:                              Zero drift ≤2% of FS/7d, Span drift ≤2% of FS/7d
Repeatability:                    Cv≤1ºC
Sample gas pressure:   0.05MPa≤Inlet pressure≤0.5MPa
Sample gas flow speed:1-3l/min
Contact capacity:              125VAC 0.3A, 110VDC 0.3A, 30VDC 1A
Signal outputs:                 4-20mA or 0-5V DC isolated
Power supply:                  220±10%VAC, 50/60Hz, power loss10VA
Working temperature:     0ºC-+45ºC
Environment humidity:    ≤90%RH
Installation dimension:  133mm*133mm*250mm
Weight:                               3.5kg
Dew Point Analyzer

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier